Product Specifications
Dimensions - 4.5"W x 15.0"H x 7.5"D
Weight - 4.5 lbs
Operating Principle - Positive Displacement
Flow Rate - 0.01-9.99 ml/hr in 0.01 ml increments10-99.9 ml/hr in 0.1 ml increments100-999 ml/hr in 1 ml increments
Volume to be Infused - 0.1-9.9 ml in 0.01 ml increments10-60 ml in 0.1 ml increments
Rate Restriction by Syringe Size:
50-60 ml syringe - 0.1-999 ml/hr30 ml syringe - 0.1-650 ml/hr20 ml syringe - 0.1-500 ml/hr10 ml syringe - 0.1-250 ml/hr5 ml syringe - 0.1-150 ml/hr3 ml syringe - 0.01-100 ml/hr1 ml syringe - 0.01-30 ml/hr
KVO Rate - 1ml/hr or selected rate, whichever is lower
Infusion Pressure Maximum - 800 mmHg without pressure sensing disc1060 mmHg with pressure sensing disc
Rate Accuracy - ±2% of full scale plunger travel not including syringe variation